Concept illustration based on manufacturer equipment. Confirm supplied components and options separately.
Instantaneous flow, totalized volume & process signals
Single-use ultrasonic flowmeter
Combine a disposable inline detector with a reusable converter to measure instantaneous flow and totalized volume in biopharma liquid processes. Review the actual flow, bubbles and tubing conditions for filtration, buffer and media preparation, or liquid transfer.
Measurement arrangement
Replaceable inline detector + reusable converter
Flow range
Model-specific ranges from 0–0.8 to 0–80 L/min
Accuracy conditions
±1% of reading at velocity ≥1 m/s · based on water calibration
Liquid operating conditions
Liquid temperature 2–60°C · pressure 0–0.5 MPa
Wetted path & sterilization
PSU · detector gamma capability up to 50 kGy
Process outputs
Configuration-dependent analog output · pulse · RS485 Modbus
Check where minimum and normal flow sit within the accuracy bands, alongside maximum capacity. Model-specific absolute errors apply at low velocity; do not interpret the entire range as ±1%.
Concept illustration based on manufacturer equipment. Confirm supplied components and options separately.
Conductive-liquid flow & replaceable flow tubes
Single-use electromagnetic flowmeter
Combine a replaceable single-use flow tube with a reusable transmitter to measure conductive-liquid flow. Review conductivity, the operating range and tube-change procedures for biopharma filtration, chromatography, and buffer or media preparation.
Model-specific ranges from 0.015–3 to 0.5–75 L/min
Accuracy conditions
Pulse: ±1% of reading with a 0.001 m/s velocity tolerance
Liquid temperature, wetted path & sterilization
Liquid temperature 2–45°C · PSU tube · C22 alloy electrodes
Process output
4–20 mA or voltage pulse
Check the liquid’s actual conductivity against the electromagnetic measurement threshold. Review cleaning, displacement or dilution steps separately when liquid conditions change within the same line.
Concept illustration based on manufacturer equipment. Confirm supplied components and options separately.
Replaceable flow path & reusable pressure gauge
Single-use pressure gauge
Measure pressure transmitted through a diaphragm from a single-use flow path to a reusable, non-wetted gauge. Review operating pressure, replacement boundaries and control signals for tubing or bag assemblies and pressure monitoring before and after filters.
Measurement arrangement
Single-use flow path + reusable non-wetted gauge
Rated measuring range
0–500 kPa
Accuracy
±3% F.S. · ±15 kPa over this range
Flow path & liquid temperature
Barb connections · size availability subject to confirmation · liquid temperature 2–60°C
Flow-path sterilization capability
Gamma up to 50 kGy · autoclave · EOG
Direct output & option
Direct 1–5 V DC · optional indicator 4–20 mA DC
Distinguish the wetted flow path from the reusable gauge when planning assembly, sterilization and disposal. Review the all-resin tube construction, which contains no metal or circuit boards, alongside management of the reusable component.
The three supplied catalogues do not establish compliance with 21 CFR Part 11. The FDA 21 CFR Part 177 reference in the electromagnetic flowmeter catalogue concerns materials, not Part 11 electronic-record and electronic-signature compliance.
A sensor output alone does not establish record-system compliance. Validation, access controls, audit trails and the use of electronic signatures need separate assessment for the intended use and complete recording system.
Illustrations show typical installation contexts, not product photos or drawings. Confirm connections and dimensions against the selected equipment's documentation.
Flow and pressure need separate measurement decisions
Relevant manufacturer: TOKYO KEISO CO., LTD.
The problem on site
If it is unclear whether a closed transfer needs instantaneous flow, batch total or pressure, the wrong sensing method can be selected. Confusing disposable and reusable components changes both the supply scope and changeover cost.
Proposed solution
Compare ultrasonic flow, electromagnetic flow and diaphragm pressure measurement by duty. Use actual liquid, conductivity, bubbles/full-pipe status, interfaces, reusable components and signal needs to narrow the options.
What to verify before selection
Check minimum/normal/maximum conditions, actual tubing, sterilisation and replacement scope, controls and record needs. Instrument output alone does not establish electronic-record compliance.
Visitor analytics
May we use Google Analytics, Google Ads and ChatGPT Ads for visitor and advertising measurement? Learn more