Email us
Conductive-liquid flow & replaceable flow tubes

Single-use electromagnetic flowmeter

Combine a replaceable single-use flow tube with a reusable transmitter to measure conductive-liquid flow. Review conductivity, the operating range and tube-change procedures for biopharma filtration, chromatography, and buffer or media preparation.

Application review requires liquid conductivity of at least 20 μS/cm; suitability does not extend automatically to low-conductivity liquids. Assess the selected tube’s temperature, pressure and materials, grounding, mounting orientation and output choice.

Request a review or quotation
Single-use electromagnetic flowmeter — equipment form illustration
An illustration based on the equipment form, not a photograph of supplied equipment or an installation drawing.

Read the solution guide

QUESTIONS THAT CHANGE THE SELECTION

Read the specification
in your operating conditions.

01

Make conductivity the first selection check.

Check the liquid’s actual conductivity against the electromagnetic measurement threshold. Review cleaning, displacement or dilution steps separately when liquid conditions change within the same line.

02

Plan the process checks around tube changes.

The documentation describes a configuration requiring no field readjustment after tube replacement. Define tube selection, mounting, grounding and signal checks alongside process-specific calibration and qualification procedures.

03

Choose the signal you need to record.

Current and pulse outputs are alternatives, not simultaneous outputs. Select the output against PLC inputs, totalization, signal resolution and the purpose of the record.

APPLICATION CONTEXT

Where will the measurement be used?

The illustration below shows an example process context for sampling and sensors. Review instrument location, a full flow path and drainability, pressure loss, cable routing and controls against the actual piping or tubing assembly.

APPLICATION IN CONTEXT
Sensors and a sampling branch are shown on a closed process vessel: signals go to the controller, while sample liquid goes to a separate collection container. 1. Process vessel; 2. Process sensor; 3. Closed sampling; 4. Analysis area; 5. Control/record
Sensors and a sampling branch are shown on a closed process vessel: signals go to the controller, while sample liquid goes to a separate collection container.
  1. Process vessel
  2. Process sensor
  3. Closed sampling
  4. Analysis area
  5. Control/record

Sample-liquid paths and sensor signals are distinct. Record measurement location, calibration, sampling time and analysis delay.

A conceptual illustration created to explain equipment and application locations. It is not an actual product photograph, installation photograph or piping/wiring design.

DATA & ELECTRONIC RECORDS

Review the measurement signal
and the electronic record.

The reviewed catalogs do not establish compliance with 21 CFR Part 11 or document electronic-record and electronic-signature controls. Analog outputs or a communications connection do not establish compliance. Review electronic-record requirements against the connected system and intended use.

Define the record

Establish whether the readings are process indicators or electronic batch records subject to regulatory recordkeeping requirements.

Review system controls

Review access, timestamps, change history/audit trails, retention, retrieval and record-copy functions in the acquisition and storage system.

Verify intended use

Review validation evidence, operating procedures and responsibilities for the intended use, including electronic-signature requirements when signatures are used.

Determine Part 11 applicability from how records required under FDA regulations are used electronically. Assess instrument documentation and the record system’s controls and validation evidence separately.

Review references: FDA · Part 11 Scope and Application21 CFR 11.10

PREPARE YOUR APPLICATION REVIEW

Start the review
with your operating conditions.

  1. 01Confirm conductivity and flow
  2. 02Compare flow tubes and wetted materials
  3. 03Review replacement and mounting procedures
  4. 04Confirm output and recording requirements
Enquire about this product and duty

Other measurement methods and process guides

INSTALLATION CONTEXT

Electromagnetic flowmeter on a transfer line

An electromagnetic flowmeter with a clear disposable flow path and external sensor head connected to transfer tubing
Electromagnetic flowmeter on a transfer line

Illustrations show typical installation contexts, not product photos or drawings. Confirm connections and dimensions against the selected equipment's documentation.

The flow range can fit while conductivity and full-pipe conditions do not

Relevant manufacturer: TOKYO KEISO CO., LTD.

The problem on site
An electromagnetic meter used for a conductive medium may not read a different buffer as reliably. Air pockets or a partly filled tube can change the recorded batch total even when the stated range looks right.
Proposed solution
Compare the disposable flow tube with actual liquid conductivity, operating flows and whether the intended location runs full. Separate disposable and reusable supply scope and specify instantaneous versus totalised values.
What to verify before selection
Check minimum conductivity against manufacturer specifications and actual liquids. Verify bubbles, full-pipe conditions, dimensions, materials, sterilisation documents and control inputs. A water calibration is not process accuracy.

Next stepCompare measurement methods and operating conditions