Make conductivity the first selection check.
Check the liquid’s actual conductivity against the electromagnetic measurement threshold. Review cleaning, displacement or dilution steps separately when liquid conditions change within the same line.
Combine a replaceable single-use flow tube with a reusable transmitter to measure conductive-liquid flow. Review conductivity, the operating range and tube-change procedures for biopharma filtration, chromatography, and buffer or media preparation.
Application review requires liquid conductivity of at least 20 μS/cm; suitability does not extend automatically to low-conductivity liquids. Assess the selected tube’s temperature, pressure and materials, grounding, mounting orientation and output choice.
Request a review or quotation
Check the liquid’s actual conductivity against the electromagnetic measurement threshold. Review cleaning, displacement or dilution steps separately when liquid conditions change within the same line.
The documentation describes a configuration requiring no field readjustment after tube replacement. Define tube selection, mounting, grounding and signal checks alongside process-specific calibration and qualification procedures.
Current and pulse outputs are alternatives, not simultaneous outputs. Select the output against PLC inputs, totalization, signal resolution and the purpose of the record.
The illustration below shows an example process context for sampling and sensors. Review instrument location, a full flow path and drainability, pressure loss, cable routing and controls against the actual piping or tubing assembly.

Sample-liquid paths and sensor signals are distinct. Record measurement location, calibration, sampling time and analysis delay.
A conceptual illustration created to explain equipment and application locations. It is not an actual product photograph, installation photograph or piping/wiring design.
The reviewed catalogs do not establish compliance with 21 CFR Part 11 or document electronic-record and electronic-signature controls. Analog outputs or a communications connection do not establish compliance. Review electronic-record requirements against the connected system and intended use.
Establish whether the readings are process indicators or electronic batch records subject to regulatory recordkeeping requirements.
Review access, timestamps, change history/audit trails, retention, retrieval and record-copy functions in the acquisition and storage system.
Review validation evidence, operating procedures and responsibilities for the intended use, including electronic-signature requirements when signatures are used.
Determine Part 11 applicability from how records required under FDA regulations are used electronically. Assess instrument documentation and the record system’s controls and validation evidence separately.
Review references: FDA · Part 11 Scope and Application21 CFR 11.10

Illustrations show typical installation contexts, not product photos or drawings. Confirm connections and dimensions against the selected equipment's documentation.