Measure pressure transmitted through a diaphragm from a single-use flow path to a reusable, non-wetted gauge. Review operating pressure, replacement boundaries and control signals for tubing or bag assemblies and pressure monitoring before and after filters.
Separate the sterilization and replacement boundaries of the disposable resin flow path and reusable gauge. Compare ±3% F.S. over 0–500 kPa with actual operating pressure; this is not a dedicated fine-differential-pressure instrument or filter-integrity tester.
Read the specification in your operating conditions.
01
Make the batch-change boundary explicit.
Distinguish the wetted flow path from the reusable gauge when planning assembly, sterilization and disposal. Review the all-resin tube construction, which contains no metal or circuit boards, alongside management of the reusable component.
02
Compare the error with the pressure change of interest.
±3% F.S. equals ±15 kPa over the 0–500 kPa range. When interpreting small changes from pressures before and after a filter, review measurement locations, errors at both points and operating flow together.
03
Check the input even without a dedicated converter.
The gauge directly outputs 1–5 V DC. Verify the PLC input, power and common-reference requirements; review the optional indicator when 4–20 mA or upper/lower alarms are needed.
APPLICATION CONTEXT
Where will the measurement be used?
The illustration below shows an example process context for sampling and sensors. Review instrument location, a full flow path and drainability, pressure loss, cable routing and controls against the actual piping or tubing assembly.
APPLICATION IN CONTEXT
12345
Sensors and a sampling branch are shown on a closed process vessel: signals go to the controller, while sample liquid goes to a separate collection container.
01Process vessel
02Process sensor
03Closed sampling
04Analysis area
05Control/record
Sample-liquid paths and sensor signals are distinct. Record measurement location, calibration, sampling time and analysis delay.
A conceptual illustration created to explain equipment and application locations. It is not an actual product photograph, installation photograph or piping/wiring design.
DATA & ELECTRONIC RECORDS
Review the measurement signal and the electronic record.
The reviewed catalogs do not establish compliance with 21 CFR Part 11 or document electronic-record and electronic-signature controls. Analog outputs or a communications connection do not establish compliance. Review electronic-record requirements against the connected system and intended use.
Define the record
Establish whether the readings are process indicators or electronic batch records subject to regulatory recordkeeping requirements.
Review system controls
Review access, timestamps, change history/audit trails, retention, retrieval and record-copy functions in the acquisition and storage system.
Verify intended use
Review validation evidence, operating procedures and responsibilities for the intended use, including electronic-signature requirements when signatures are used.
Determine Part 11 applicability from how records required under FDA regulations are used electronically. Assess instrument documentation and the record system’s controls and validation evidence separately.
Illustrations show typical installation contexts, not product photos or drawings. Confirm connections and dimensions against the selected equipment's documentation.
Transfer pressure is not the same measurement as a small pressure differential
Relevant manufacturer: TOKYO KEISO CO., LTD.
The problem on site
Watching for a blocked path or a large pressure swing requires different range and resolution from measuring a small difference across a filter. The reusable gauge's signal also may not match an existing controller.
Proposed solution
Review the replaceable wetted tube and reusable diaphragm-coupled gauge separately. Compare the operating window, transients, acceptable error and input signal, treating small differential pressure and integrity testing as different duties.
What to verify before selection
Confirm minimum/normal/maximum pressure, the smallest change of interest, interfaces, contact materials and sterilisation documentation, plus actual outputs and receiving inputs. Analogue output alone does not establish electronic-record suitability.